Over Current Relay Testing System Manufacturer,Supplier and Exporter in India

Product Code : SCL-EELE-12977

Indian Lab Suppliers supplies the Over Current Relay Testing System (Product ID: 12977) as part of its electronics engineering lab equipment range. It is a single-phase test panel for plotting the time-current characteristic of an electromechanical inverse-time overcurrent relay.

Quick Answer

The relay works from a 1 A CT secondary and has seven plug settings from 0.5 A to 2 A in 0.25 A steps. Current is raised through a variac (0–270 V, 0–5 A) and a 110 Ω, 5 A rheostat, and a timer (10 ms to 30 min) starts with the injected current and stops when the relay trips, so the operating time is recorded automatically. Repeating this at several multiples of the plug setting gives the inverse curve: the larger the current, the faster the trip. Voltage and current are shown together on a 16 × 2 LCD, and a diagram shows where the relay sits on a transmission line.

Key Specifications

ParameterDetail
Mains supply230 V ±10%, 50 Hz
RelayElectromechanical, inverse time, 110 V AC 50 Hz, CT secondary 1 A
Current settings0.5, 0.75, 1, 1.25, 1.5, 1.75, 2 A
VariacInput 230 V; output 0–270 V; 0–5 A
Rheostat110 Ω, 5 A
Voltmeter25–300 V
Ammeter200 mA–5 A
Timer10 ms–30 min, automatic start/stop with the relay
IncludedLearning material CD; 2-year warranty

Where This Is Used

Used in switchgear and protection laboratories for relay setting and grading exercises. Terminals are at mains voltage and the rheostat gets hot at high current, so connect with the supply off. Related: IDMT Over Current Relay Kit (Product ID: 12995) and IDMT And Directional Over Current Relay Setup (Product ID: 10405). Browse the full Electronics Engineering Lab Equipment range or related Electricity and Magnetism products.

Frequently Asked Questions

What current settings does the relay have?
0.5 A to 2 A in 0.25 A steps, on a 1 A CT secondary.

How is tripping time measured?
By a timer that starts with the current and stops when the relay operates.

Is the relay electromechanical?
Yes, so its internal mechanism can be studied as it operates.