Fuse, MCB & Bimetallic Relay Characteristics Trainer Manufacturer,Supplier and Exporter in India

Fuse, MCB & Bimetallic Relay Characteristics Trainer

Product Code : SCL-EELE-12854

Indian Lab Suppliers supplies the Fuse, MCB & Bimetallic Relay Characteristics Trainer (Product ID: 12854) as part of its electronics engineering lab equipment range. It is a protection trainer for plotting the time-current characteristics of a fuse, an MCB and a bimetallic overload relay.

Quick Answer

A current injector, built from a 230 V / 1 A dimmer and a short-circuit transformer with 0-2-8 V / 12 A secondary taps, drives a chosen current through the 2 A rewireable fuse, 2 A MCB or 2 A bimetallic relay, while a 20 A AC ammeter and an elapsed-time counter with 1 ms resolution record the trip. A Hall-sensor zero-current detector panel (up to 20 A, 2.1 kV isolation) with a DC voltage transducer and relay contacts stops the timer. A control supply panel feeds DC to the signal panels and distributes single-phase AC through three 4 A MCBs, all on an aluminium profile panel system with 4 mm shrouded sockets. An air circuit breaker (400 A, 415 V) and an 11 kV SF6 breaker are offered as options for high-voltage labs; please confirm them at quotation, and apply test currents only under supervision.

Key Specifications

ParameterDetail
Devices testedRewireable fuse 2 A, MCB 2 A, bimetallic relay 2 A
Current injector1-phase dimmer 230 V/1 A; short-circuit transformer 230 V/1 A primary, 0-2-8 V/12 A taps
Timing and meteringAC ammeter 20 A; elapsed-time counter, 1 ms resolution
Zero-current detectorHall-sensor current measurement up to 20 A, 2.1 kV isolation, 0-2.5 V output; DC voltage transducer up to 600 V; relay NO/NC contact
Control supplyDC multi-output supply; 3 × 4 A MCBs for single-phase AC distribution
Optional breakersACB 3-pole 415 V, 400 A, 50 kA; SF6 breaker 11 kV, 630 A
Dimensions1165 (L) × 300 (W) × 545 (H) mm
DocumentationInstructor guide and student workbook

Where This Is Used

Used in switchgear and protection labs, before relay trainers such as the Percentage Biased Differential Relay Trainer (Electromech / Numeric) (Product ID: 12855); basic earthing and leakage work is covered by the Safety and Earthing Module (Product ID: 12844). Related items: MCB and HRC Fuse Testing System (Product ID: 12976). Browse the full Electronics Engineering Lab Equipment range or related Electricity and Magnetism products.

Frequently Asked Questions

Which characteristics are plotted?
Time-current curves of a fuse, an MCB and a bimetallic relay.

How is tripping time measured?
With an elapsed-time counter of 1 ms resolution.

Are ACB and SF6 breakers included?
They are optional; please confirm at quotation.