Combined Earth Fault and Over Current Numerical Relay Kit Manufacturer,Supplier and Exporter in India

Product Code : SCL-EELE-12996

Indian Lab Suppliers supplies the Combined Earth Fault and Over Current Numerical Relay Kit (Product ID: 12996) as part of its electronics engineering lab equipment range. It is a rack-mounted kit for testing a numerical relay that combines IDMT overcurrent and earth fault protection.

Quick Answer

Test current is set with a single-phase dimmer (230 V, 1 A) feeding a short-circuit transformer (230 V, 1 A primary; 0–2–8 V, 12 A secondary taps), which gives the high current at low voltage needed to operate the relay. An over-current and elapsed-time panel with a 20 A ammeter records the trip time up to 999 s at 1 ms resolution, and all relay connections, including 2 NO trip contacts, are brought out on the relay panel. The relay is rated 5 A with an adjustable current setting. Two experiments are listed: IDMT overcurrent protection and earth fault protection. Please confirm the relay setting ranges at quotation.

Key Specifications

ParameterDetail
Input panelSingle phase AC; MCBs 4 A and 1.6 A; bulb load
Injector1-phase dimmer 230 V / 1 A; short-circuit transformer 230 V / 1 A primary, 0–2–8 V / 12 A secondary
Timer panelAC ammeter 20 A; elapsed time up to 999 s, 1 ms resolution
RelayNumerical IDMT overcurrent with earth fault, 5 A, adjustable current setting
Relay panelAll connections brought out; 2 NO trip contacts
FrameAluminium profile rack with ABS panels and 4 mm shrouded sockets

Where This Is Used

Used in protection laboratories for overcurrent and earth fault relay tests. The injector secondary carries up to 12 A and the transformer and leads warm up, so keep test runs short and connect with the supply off. Related: Three Phase Over Current & Earth Fault Numeric Relay Testing System (Product ID: 12985) and Earth Fault Relay Testing System (Product ID: 12979). Browse the full Electronics Engineering Lab Equipment range or related Electricity and Magnetism products.

Frequently Asked Questions

How is high test current produced?
With a short-circuit transformer giving up to 12 A at 2 or 8 V.

What is the relay current rating?
5 A.

What is the timer resolution?
1 ms.

Related Categories

Electronics Engineering Lab Equipment · Electricity and Magnetism · Engineering Lab Equipment

Request a Quote

For pricing, bulk quantities or tender documentation on the Combined Earth Fault and Over Current Numerical Relay Kit, reach our team via Contact Us or submit your requirement through our Tenders page.

   

Office Address

Works: #975 Malleshwaram,
Bengaluru, Karnataka 560003

[email protected]

Follow Us